Ensuring System Wellness (IC Devices)
Uploading ICS-1500/2000 or ICS-1600/2100 Calibration Data
After performing a calibration procedure or changing a calibration variable from the ICS-1500/2000 or ICS-1600/2100 touch screen, use the procedure below to update the System Wellness database with the new calibration data.
Note:
Do not open a System Wellness Control Panel if the name includes "Service.pan." These Wellness panels are reserved for use by Dionex Service Representatives.
Open the System Wellness Control Panel for the ICS-1500/2000 or ICS-1600/2100 (see Opening a Wellness Control Panel).
Locate the Update Wellness Database controls. If desired, click Instructions for an overview of the procedure.
Click Upload to begin uploading (sending) calibration data to Chromeleon.