Ensuring System Wellness (IC Devices)
Calibrating the ICS-900 Pressure Transducer

 

When to Calibrate:

Every 6 months

 

  1. Open the ICS-900 System Wellness Control Panel (see  Opening a Wellness Control Panel).

  2. Under Pressure Transducer, click the Calibrate Offset button. A panel with instructions and command buttons will open.

  3. Follow the instructions on the panel to complete the transducer offset calibration procedure. The new offset is determined and stored as the current value.

  4. Close the waste valve.

  5. Under Pressure Transducer, click the Calibrate Slope button. A panel with instructions and command buttons will open.

  6. Follow the instructions on the panel to complete the transducer slope calibration procedure. The new slope is determined and stored as the current value.

  7. Turn off the pump.

  8. Disconnect the pressure gauge and backpressure coil.

  9. Reconnect the pressure transducer to the pump.