Ensuring System Wellness (IC Devices)
Calibrating the Leak Detector

When to Calibrate:

After installing a new leak detector

If the leak detector diagnostic test fails

Every 6 months

 

Many System Wellness supported devices are equipped with leak detectors. The calibration procedure for all of the sensors of IC devices is the same.

  1. Thoroughly dry the sensor.

  2. Open the System Wellness Control Panel for the device (see  Opening a Wellness Control Panel.)

  3. Under Calibration, leak detector, click the internal script button to calibrate the sensor installed in the device itself. Click the external script button to calibrate the sensor on a controlled device. For example, for a pump, internal refers to the sensor in the pump and external refers to the sensor in a chromatography oven controlled by the pump.

  4. The device calibrates the sensor and uploads the new value to Chromeleon. Chromeleon stores this new calibration value as the current value.