Ensuring System Wellness (IC Devices)
Calibrating the ICS-900 Flow Rate

 

When to Calibrate:

  • If the OQ/PQ flow rate accuracy and precision test fails

 

 

Items Needed:

  • 0.076-mm (0.003-in) ID yellow PEEK tubing (P/N 049715) to create 14 ± 2 MPa (2000 ± 300 psi) of backpressure (if needed)

  • High purity ASTM Type I (18 megohm-cm) filtered and deionized water

  • Balance capable of weighing more than 10 g with 0.001 g readability

  • Tared beaker

 

  1. Open the System Wellness Control Panel for the ICS-900 (see  Opening a Wellness Control Panel).

  2. Under Pump Flow Rate Calibration controls, click Calibration Procedure. A panel with instructions and command buttons will open.

  3. Follow the instructions on the panel to complete the calibration procedure. The new calibration value is determined and stored as the current value.

  4. When calibration is complete, click Log to record the new calibration value in the Audit Trail.

Related Topics

  Performing Device Calibrations