Ensuring System Wellness (IC Devices)
Calibrating the ICS-1000/1500/2000 or ICS-1100/1600/2100 Degas Pump

When to Calibrate:

Every 6 months

 

 Note:

Do not open a System Wellness Control Panel if the name includes "Service.pan." These Wellness panels are reserved for use by Dionex Service Representatives.

  1. Open the System Wellness Control Panel for the ICS-1000/1500/2000 or ICS-1100/1600/2100 (see  Opening a Wellness Control Panel).

  2. Locate the Degas Calibration controls. If desired, click Instructions for an overview of the calibration procedure.

  3. Click Calibrate to begin the calibration. When the calibration is complete, the ICS-1000/1500/2000 or ICS-1100/1600/2100 will upload the new calibration value to Chromeleon. Chromeleon will store the new calibration value as the current calibration value.

  4. Click Log to record the new calibration value and degas pressure in the Audit Trail. Chromeleon will automatically update the Current Cal. Date and Last Calibration Date fields.