Appendix: Cables and Pin Assignments
Agilent Remote Cable (LC System)

Instrument for which the cable is required:

External autosamplers (e.g., CTC PAL or UltiMate 3000 autosamplers) that are operated with Agilent LC systems controlled by the Instrument Control Framework.

Cable type:

Connector 01046-60201 (general purpose cable). The cable is provided by Agilent.

 

Pin assignment:

Pin

Universal

Pin

Agilent 1100

Signal Name

Active

(TTL)

 

1 -  White

Digital ground

 

 

2 - Brown

Prepare run

Low

 

3 - Gray

Start

Low

 

4 - Blue

Shut down

Low

 

5 - Pink

Not connected

 

 

6 - Yellow

Power on

High

 

7 - Red

Ready

High

 

8  - Green

Stop

Low

 

9 - Black

Start request

Low

 

To start, short wire 1 and 3; to shut down (load error method), short wire 1 and 4.

 

Connecting an UltiMate 3000 Autosampler to the LC System

In order to operate a WPS-3000 autosampler with the Agilent LC system, you also need the following cable:

6-pin Mini-DIN signal cable (5m; Dionex part no. 6000.1004)

 

To connect the signal cable with the remote cable of the LC system:

  1. Connect one end of the signal cable to relay 4 of the digital I/O port of the autosampler.

  2. Connect the Agilent general purpose remote cable with the LC system.

  3. Connect the remote cable with the other end of the Dionex signal cable: connect wire 1 of the Agilent remote cable to the brown wire, and wire 3 to the white wire of the Dionex signal cable.

 