Installing and Controlling Third-Party Devices
Rheodyne MXII Series Valves
Device Type: |
Switching valve |
Device Driver: |
Rheodyne: MXII Valve The driver was tested with a 10-position, 11-port valve using valve controller firmware revision 41. |
Hardware Options |
|
What is required?
License: |
Timebase Class 1 |
Connection: |
The following items are required to connect the valve to the Chromeleon Server PC:
|
To control the instrument, use the Rheodyne MXII Series.pan panel from the |
Hardware Installation
Device Connection: |
|
Device Settings: |
The following settings are supported: Baud rate 19200 Data bits 8 Stop bits 1 Parity None Handshake Off Tip: For additional information on USB control, refer to the MX Series II Quick Start Guide 2321507B provided with the valve. |
Software Installation
For details about how to add timebases and device drivers to the Server Configuration, refer to Adding, Configuring or Deleting Devices. When you add a driver to the timebase, the configuration dialog or wizard opens automatically. For information about the settings, click the Help button on a page or press F1.
Troubleshooting
The following error may occur:
After switching the valve into position 2, the valve changes back to position 1. This is followed by the subsequent error message:
{Valve} A timeout occurred while waiting for the device to reach specified position in TaskSetPosition
and the valve is disconnected.
Remedy: Power-cycle the device and reconnect once with Chromeleon.
Further Information
For detailed installation instructions, refer to the operating instructions for the valve.
For an overview of manufacturers whose instruments can be controlled by Chromeleon, refer to Installing and Controlling Third-Party Devices.