Installing and Controlling Third-Party Devices
Gilson INI File (Example)
The following example describes the structure of an INI File typical for Gilson Autosampler. Select the INI File on the INI File tab page. See the Comment section for a description of the individual lines.
Section |
Comment |
[Gilson-231] |
|
SamplerID = 10 |
GSIOC sampler address |
DilutorID = 0 |
GSIOC address |
Syringe = 500 |
Dilutor syringe volume |
RackCode = 0 |
Rack code (to identify rack) |
TriggerInject = 1 |
Determines which change from load to inject position will be interpreted as inject response |
|
|
[Inject] |
|
1 INJECT 1 |
Rheodyne in inject position |
2 RACK CODE 0 |
Rack code = 0 |
3 C0 = 1 |
Number of samples = 1 |
4 INPUT C1/1 |
Start position enquiry |
5 IF C0 + C1 > 81 |
Safety condition |
6 GOTO 3 |
|
7 INPUT C2/31 |
Get inject volume |
8 B3 = 1 |
B3 = variable for y coordinate |
9 FOR A = 1/5 |
Calculus of Y-(B3) and |
10 B = C1 - 16 |
X-(C1) coordinates of sample position |
11 IF B < 1 |
|
12 GO TO 16 |
|
13 C1 = C1 - 16 |
|
14 B3 = B3 + 1 |
|
15 NEXT A |
|
16 C9 =1 |
Variable for program start |
17 C = 0 |
Set sample counter to 0 |
18 RINSE |
Wash syringe needle outside |
19 DISP. 0/500/9 |
|
20 TUBE 0/0 |
Wash needle inside |
21 DISP. 0/500/4 |
Wash inject port |
22 FOR B = B3/5 |
Y- direction on rack |
23 FOR A = C1/16 |
X- direction on rack |
24 C = C + 1 |
Increment sample counter |
25 PRINT C/1 |
Display sample number |
26 IF C > C0 |
Reset condition |
27 HOME |
|
28 TUBE A/B |
Go to position A/B |
29 HEIGHT |
Move needle to maximum elevation |
30 WAIT 2 |
Wait 2/100 min. |
31 IF C9 = 1 |
Variable for program start |
32 GO TO 34 |
|
33 WAIT /3/0 |
Wait for address 3 to reach state 0 |
34 ASPIR 0/5/1 |
Draw 5 µl air |
35 HEIGHT 0 |
Move needle to lowest position |
36 ASPIR 0/C2 + 15/1 |
Draw inject volume + 15 µl |
37 WAIT 2 |
Wait 2/100 min. |
38 TUBE 0/0 |
Inject position |
39 DISP 0/15/1 |
Wash dead volume with 15 µl sample solvent |
40 WAIT 2 |
Wait 2/100 min. |
41 INJECT 0 |
Move Rheodyne valve to load position |
42 DISP 0/C2/1 |
Load sample volume to loop |
43 WAIT 2 |
Wait 2/100 min. |
44 INJECT 1 |
Rheodyne to inject position |
45 AUXIL 7/1 |
Start integrator |
; 45 AUXIL 1/1 |
Command for device no. 203 994 and 203 995 |
46 WAIT 5 |
5/100 min. waiting time |
47 AUXIL 7/0 |
Reset integrator start signal |
; 47 AUXIL 1/1 |
Command for device No. 203 994 and 203 995 |
48 C9 = 2 |
Program continuation variable |
49 DISP 0/500/9 |
Wash |
50 RINSE |
|
51 DISP 0/500/9 |
|
52 NEXT A |
Next sample in X- direction |
53 C1 = 1 |
Set X coordinate for next sample row |
54 NEXT B |
Next row |
56 HOME |
|
*** END OF INJECT PROGRAM |
|
|
|
[Prog-0] |
|
[Prog-1] |
|
[Prog-2] |
|
[Prog-3] |
|
[Prog-4] |
|
[Prog-5] |
|
[Prog-6] |
|
[Prog-7] |
|
[Prog-8] |
|
[Prog-9] |
|
For an overview of the different Gilson instruments for which device drivers are available in Chromeleon, refer to Gilson.