Installing and Controlling Dionex Devices
External MSV: Installation
Device Type: |
External motorized switching valve (MSV) |
Device Driver: |
Dionex: External MSV |
Hardware Options |
|
What is required?
License: |
Timebase Class 1
|
Connection: |
Use the modem cable (Sub-D 9 female to Sub-D 9 female) shipped with the valve. Note: You cannot use the standard Dionex modem cable in this case. |
Hardware |
A free COM port is required for each single valve. |
Hardware Installation
Device Connection: |
Use the modem cable (Sub-D 9 female to Sub-D 9 female) shipped with the device to connect the valve. Note: You cannot use the standard Dionex modem cable in this case. |
Device Settings: |
The following settings are supported: Baud rate 9600 Data bits 8 Stop bits 1 Parity None Handshake Off
The DIP switches are set as follows: SW3-1 ON SW3-2 ON SW3-3 ON: RS-232 control via J2 SW3-4 ON: 9600 Baud SW3-5 ON SW3-6 ON SW3-7 ON SW3-8 OFF |
Software Installation
For details about how to add timebases and device drivers to the Server Configuration, refer to Adding, Configuring or Deleting Devices. When you add a driver to the timebase, the configuration dialog or wizard opens automatically. For information about the settings, click the Help button on a page or press F1.
Troubleshooting
If communication problems occur, verify that the DIP switches are set correctly (see Device Settings above).
Further Information
Tip:
For more installation details, refer to the manual for the instrument.
For an overview of the different Dionex devices that are distributed at present, refer to Installing Dionex Devices.