Installing and Controlling Dionex Devices
 External MSV: Installation
 External MSV: Installation
| Device Type: | External motorized switching valve (MSV) | 
| Device Driver: | Dionex: External MSV | 
| Hardware Options | 
 | 
What is required?
| License: | Timebase Class 1 
 | 
| Connection: | Use the modem cable (Sub-D 9 female to Sub-D 9 female) shipped with the valve. 
 You cannot use the standard Dionex modem cable in this case. | 
| Hardware | A free COM port is required for each single valve. | 
Hardware Installation
| Device Connection: | Use the modem cable (Sub-D 9 female to Sub-D 9 female) shipped with the device to connect the valve. 
 You cannot use the standard Dionex modem cable in this case. | 
| Device Settings: | The following settings are supported: Baud rate 9600 Data bits 8 Stop bits 1 Parity None Handshake Off 
 The DIP switches are set as follows: SW3-1 ON SW3-2 ON SW3-3 ON: RS-232 control via J2 SW3-4 ON: 9600 Baud SW3-5 ON SW3-6 ON SW3-7 ON SW3-8 OFF | 
Software Installation
For details about how to add timebases and device drivers to the Server Configuration, refer to  Adding, Configuring or Deleting Devices. When you add a driver to the timebase, the configuration dialog or wizard opens automatically. For information about the settings, click the Help button on a page or press F1.
 Adding, Configuring or Deleting Devices. When you add a driver to the timebase, the configuration dialog or wizard opens automatically. For information about the settings, click the Help button on a page or press F1.
Troubleshooting
If communication problems occur, verify that the DIP switches are set correctly (see Device Settings above).
Further Information
 Tip:
 Tip:
For more installation details, refer to the manual for the instrument.
For an overview of the different Dionex devices that are distributed at present, refer to  Installing Dionex Devices.
 Installing Dionex Devices.