Tab Dialog
Box:
Sampler (6890/6850 GCs)
Also, refer to Hardware Installation Agilent 6890 GC and Agilent 6850 GC.
Automatic Liquid Sampler Installed: |
Select this check box if you work with an automatic liquid autosampler. |
|
|
Name: |
Enter the autosampler name. |
|
|
Tray
Installed |
Select this check box if a tray is installed. |
|
|
Select this check box if the installed injector supports Fast Multiple Injection and you want to use this mode. Note: Fast multiple injection is only supported for firmware version N.06.07 and above. |
|
|
|
Bar
Code Reader Installed |
Select this check box if a bar code reader is installed. Select the UseBCR = On command to have the read value entered in the Audit Trail. |
(Front/Back) Injector
Installed: |
Select this check box if you use a (front/back) injector. (GC 6850 only): Click the arrow to select the installed injector (supported injector: 7683 ALS) or the type of tray (22 vials or 27 vials). (6890 GC only): Click the arrow to select the installed injector (supported injector types: Agilent 7673 ALS, 7683 ALS and 7693 ALS). Note: If both injectors are installed on the 6890 GC, Dual Inject is automatically enabled. Both injectors can inject simultaneously from two different sample positions. |
|
|
Name: |
Enter the injector name. |
|
|
Syringe: |
Enter the currently used syringe type. |
Tip:
Errors in the configuration will not be recognized! The entered syringe type must correspond to the one that is actually used; otherwise, the calculations are based on wrong injection volumes.