Tab Dialog
Box:
Sampler (6890/6850 GCs)
Also, refer to Hardware
Installation Agilent 6890 GC and
Agilent 6850 GC.
Automatic Liquid Sampler Installed: |
Select this check box if you work with an automatic liquid autosampler. |
|
|
Name: |
Enter the autosampler name. |
|
|
Tray
Installed |
Select this check box if a tray is installed. |
|
|
Select this check box if the installed injector supports Fast Multiple Injection and you want to use this mode.
Fast multiple injection is only supported for firmware version N.06.07 and above. |
|
|
|
Bar
Code Reader Installed |
Select this check box if a bar code reader is installed. Select the UseBCR = On command to have the read value entered in the Audit Trail. |
(Front/Back) Injector
Installed: |
Select this check box if you use a (front/back) injector. (GC 6850 only): Click the arrow to select the installed injector (supported injector: 7683 ALS) or the type of tray (22 vials or 27 vials). (6890 GC only): Click the arrow to select the installed injector (supported injector types: Agilent 7673 ALS, 7683 ALS and 7693 ALS).
If both injectors are installed on the 6890 GC, Dual Inject is automatically enabled. Both injectors can inject simultaneously from two different sample positions. |
|
|
Name: |
Enter the injector name. |
|
|
Syringe: |
Enter the currently used syringe type. |
Tip:
Errors in the configuration will not be recognized! The entered syringe type must correspond to the one that is actually used; otherwise, the calculations are based on wrong injection volumes.