Tab Dialog Box:  
Options (1100/1200 HPLC System)

To enable control of the optional components in 1100 or 1200 HPLC System devices, select the corresponding check boxes. Only selected components can be controlled.

 

Installed Options for Pump

Select the Solvent Selection Valves check box if solvent selection valves are installed (binary pumps only).

Select the Rear Seal Wash check box if the rear seal wash option is installed.

If the Communication error: LICOP command not acknowledged. (RE0503 command unknown (outdated firmware?)) message appears after installation, select the Disable Solvent Limit Features check box. If this message does not appear, Dionex recommends accepting the default setting, i.e., the check box is cleared. This setting allows you to use the EmptySolventError and SolventNotReadyLimit commands for  Checking the Solvent Liquid Level.

Select the External Contact Board check box if an external relay contacts board is installed in the pump.

 

 

Installed Options for Sampler

Select the External Contact Board check box if an external relay contacts board is installed in the autosampler.

(Agilent 1200 HPLC system only) Select the Disable Tray Light Feature check box if the installed autosampler does not support tray lighting.

 

 

Installed Options for Column Compartment

Select the Column Switching Valve check box if a column switching valve is installed.

 

 

Installed UV Lamp for VWD Detector

Select P/N 2140-0590 (heat during ignition) if a lamp with this part number is installed. This lamp needs to be heated during ignition.

Select P/N G1314-60100 (continuous heating) if a lamp with this part number is installed. This lamp needs to be heated during ignition and operation.

 

 

Installed Options for Fraction Collector

Select the Disable Tray Light Feature check box if the installed fraction collector does not support tray lighting.

Select the External Contact Board check box if an external relay contacts board is installed in the fraction collector.

 

For more information, refer to Hardware Installation  Agilent 1100 HPLC System or  Agilent 1200 HPLC System.