Tab Dialog
Box:
Front/Back Inlet (TRACE 1300 Series GC)
Use the front and back inlet tab pages to define the inlets installed on the GC.
Option |
Description |
|
|
Installed |
This check box appears on both the front and the back inlet tab page. Select this check box if an inlet is installed in the respective position on the GC. |
|
|
Name |
Enter a name for the front or back inlet, if desired, or leave as default. |
|
|
Type |
Select the inlet type: SSL or PTV. Note: If the PTV inlet is selected, the Cryogenics Installed check box becomes available. |
Column Name |
Enter a column name, if desired, or leave as default. |
|
|
Carrier Gas |
Select the carrier gas from the drop-down list. Any changes will be sent to the GC after clicking OK. |
|
|
Maximum Temperature |
Specify the maximum inlet temperature in the available range. The range changes if the cryogenics option is installed on the GC. |
|
|
Minimum Temperature |
Specify the minimum inlet temperature in the available range. The range changes if the cryogenics option is installed on the GC. |
|
|
Maximum Pressure |
Specify the maximum pressure in the available range. |
|
|
Maximum Flow |
Specify the maximum flow in [ml/min]. |
|
|
Maximum Pressure Rate |
Specify the maximum pressure rate. Note: The Maximum Pressure Rate setting is only available if one of the following applies:
|
|
|
Select this check box if a backflush option is installed for this inlet on the GC. Backflush is used to backflush the sample out of the sample loop. |
|
|
|
Cryogenics Installed |
Select this check box if a cryogenics option is installed on the GC for this inlet. |
|
|
(SSL inlet only) Select this check box if a Helium Saver SSL Inlet is installed on the GC, which serves to control helium flow. Note: Selecting this check box disables the check boxes Backflush Option Installed and Gas Sampling Valve. |
|
|
|
(SSL inlet only) Select this check box if a Gas Sampling Valve (GSV) is installed on the GC. GSVs help to switch gas streams to be introduced into a column. Notes:
|
Get |
Click this button to download the hardware configuration from the GC. |
Note:
If demo mode is off, the Type, Backflush Option, Helium Saver, Gas Sampling Valve and Cryogenics controls are disabled but will reflect the configuration of the GC.