Maximum Width
Location: |
Detection tab page |
Type: |
Fixed point number |
Dimension: |
min |
Value Range: |
0.01 ... 999.99 |
Default: |
The presetting is 999.99 min so that all peaks are considered. |
Related Parameters: |
|
Description: |
The Maximum Width detection parameter defines the maximum width above which peaks are ignored during peak detection. The peak width is measured on the baseline. For peaks that do not reach the baseline, the width is extrapolated. The parameter must be enabled before the peak end. The maximum width influences peak recognition and thus the baseline. |
Tip: |
You can hide peaks with a small area in the report without changing peak detection. To do so, in the report, select Properties on the Table menu, and then select the Reject peaks with smaller relative area than ...% check box. |
For information about how to use individual detection parameters, refer to Integrating Chromatograms and Identifying Peaks Defining Detection Parameters.