The ESA Coulochem II Detector allows you to store 10 Pulse or DC methods. In this context, a method is a combination of various parameters for a specific application. These include:

 

- Cell potentials (E1, E2, E3)

 

- Guard cell potentials

- Pulse widths (T1, T1´, T2, and T3)

 

- Filter constants

- Voltage range

 

- Sensitivity range

- Channel offset

 

 

 

Determining and storing methods is only possible directly on the detector. Refer to the Detector manual for details.

On the ESA Control Panel:

Click Run Method to execute the method (1 to 10) included in a program line. If the parameter 0 is entered, the current; that is, the latest settings are used.

Syntax: tCouloChem.RunMethod no., mode

For example:

 0CouloChem.RunMethod 8, Pulse, or

 3CouloChem.RunMethod 4, DC

Click Stop Method to stop the active method.

Click Autozero to perform an autozero for the current channel.

Click (Scroll) Down to display the end of the data on the detector display.

Click (Scroll) Up to display previous data on the detector display.