Detector Control
Selecting a Default Waveform for Integrated Amperometry
Chromeleon provides default waveforms for the AgCl, pH, and PdH reference modes. You can use these waveforms as is, or modify them for your own applications. If you modify one of the default waveforms, you must save it with a different name. The default waveforms are read-only.
Waveform Name |
Waveform Description |
Working Electrode* |
Reference Mode** |
Comments |
Gold, Ag-AgCl RE, Carbo, Quad |
Carbohydrates, Quad Potential, Ag/AgCl Ref |
Au disposable (P/N 060139; pkg. of 6) |
AgCl |
Can be used with disposable gold electrodes. |
Gold, PdH RE, Carbo, Quad |
Carbohydrates, Quad Potential, PdH Ref |
Au disposable (P/N 060139; pkg. of 6) |
PdH |
Currently supported only for carbohydrate detection on gold electrodes (gaskets for capillary format required). To use PdH referencing for other applications, contact Technical Support for Dionex products for the latest information and/or see the Note below. |
Gold CWE, Ag-AgCl RE, Carbo, Triple |
Carbohydrates, Triple Potential, Conventional Electrode Only |
Au (P/N 044112) |
AgCl |
DO NOT use with disposable gold electrodes; dissolves gold and can cause reproducibility problems. |
Gold, Ag-AgCl RE, AAA |
Amino Acids, Carbohydrates, Water-Soluble Vitamins |
Au disposable (P/N 060082; pkg. of 6) |
AgCl |
For research tasks (e.g, waveform development or separation optimization), use non-disposable gold electrode (P/N 055832). |
Gold, pH-Ag-AgCl RE, AAA |
Amino Acids, Carbohydrates, Water-Soluble Vitamins |
Au disposable (P/N 060082; pkg. of 6) |
pH |
For research tasks (e.g, waveform development or separation optimization), use non-disposable gold electrode (P/N 055832). |
Platinum, Ag-AgCl RE, Acidic |
Alcohols, Glycols, Carboxylates, Cyanide |
Pt (P/N 044113) |
AgCl |
Use with platinum electrodes only. |
Platinum, Ag-AgCl RE, Acidic (Old) |
Alcohols, Formic Acid, Glycols |
Pt (P/N 044113) |
AgCl |
Use with platinum electrodes only. |
Silver, Ag-AgCl RE, S-2, CN-, I-, S2O3-2, Alkaline |
Sulfide, Cyanide, Iodide, Thiosulfate |
Ag disposable (P/N 063003; pkg. of 6) |
AgCl |
|
Silver, Ag-AgCl RE, Br-, SCN-, Alkaline |
Bromide, Thiocyanate |
Ag disposable (P/N 063003; pkg. of 6) |
AgCl |
|
* Working electrode only (does not include complete electrode cell). |
||||
** The same combination electrode (P/N 044198) is used for both the AgCl and pH reference modes. |
Note on Converting Ag/AgCl Waveforms for the PdH Reference Electrode
If you want to convert a Ag/AgCl-referenced waveform for use with the PdH reference electrode, add the following potential offset value to each potential of the waveform:
Potential Offset = (eluent pH x 59) + 148
Note:
To date, this calculation has been confirmed for alkaline, aqueous eluents only. Contact Technical Support for Dionex products for information regarding PdH referencing in other types of chromatographic eluents.