Commands and Tips for Third-Party Devices
Waters 717plus Autosampler

For information about the standard autosampler commands, refer to Device Control  Dionex Autosamplers.

In addition, Chromeleon supports the following properties:

 

Property

Min

Max

Description

Purge

 

Enables purging of the sample loop and injection syringe with mobile phase. Also, see PurgeTime.

PurgeTime

1.0

99.0

Specifies for how long (in min) purging of the sample loop and injection syringe shall be performed.

SyringeVol

25, 250, 2500

Indicates the syringe volume in µl (read-only).

InjectWaitTime

0

1E+7

Sets the time between the inject command and the inject response in seconds. The default setting is 0 s.

LoopPressure

-3,45

380

Indicates the pressure (in bar) in the sample loop (read-only). This property is available only on Expert level.

 

 Note:

As described in the instrument's manual, negative pressure readings are possible for the sample loop pressure (LoopPressure property).

 Tip:

Make sure that filled vials are available on the positions defined in the sequence. Since the autosampler does not recognize if the vials are empty or missing, injections are performed from empty positions, also.

For installation instructions, refer to Hardware Installation  Waters 717plus Autosampler: Installation in the Administrator Help section.