Commands and Tips for Third-Party Devices
Varian 3400 and 3600 GCs: Troubleshooting
Problem: The GC does not reach the Ready state
In addition, the following message appears in Chromeleon: GC.CausesForNotReady=Remote_In.
Solution:
Disable the Remote input on the device. Select Set Checks on the Configuration menu and set Wait for ext. device ready? to No.
Error messages in Chromeleon:
(GC) Communication Error: Bad Response to "0" command 0
Problem: The instrument uses a method from the GC's method table.
The error message is repeated until Chromeleon disconnects from the instrument.
Solution: On the instrument, delete all references to Method 1:
Select Method 1.
Press Delete/Section Table and select Section or Table, and then select Method 1. Select the section you want to delete (here: Column). The following message should appear on the instrument: Section Cleared.
Repeat these steps for the Injection Port, Autosampler, and Detector sections of the instrument.
If you want to use a Varian GC method (instrument) in Chromeleon, you have to add an autosampler. Select Build/Modify, and then select Method 1. Enable the autosampler in the method by selecting the autosampler and adding it to Method 1.
(GC) Communication Error: Bad Response to XB?22@DS Command: XB?22@DS COMMAND (or … to XC?22@DX Command: XC?22@DS COMMAND)
Problem: No autosampler has been defined in the Varian GC method (on the instrument).
Solution: Add the autosampler to the Varian GC method in Single mode. Select the autosampler on the instrument's panel via Build/Modify > Method 1 > Autosampler > Add Autosampler Section > Yes.
(GC) Communication Error: Bad Response to "Read Flag Not Ready" Command: ;B?22@DS COMMAND (or C?22@DS COMMAND)
Problem: Communication error due to incompatible EPROM or Varian GC CPU board.
Solution: Replace the four outdated EPROMs with EPROMs version 03-910617-00 Rev. 1 or higher.
Autosampler not present:
Problem: Communication error with the autosampler, mainly due to a bad connection.
Solution: Disconnect the autosampler from the power source and connect the brown autosampler cable to the CPU board.