Commands and Tips for Third-Party Devices
Shimadzu GC-2010/2014 GC System: Autosampler
MainInjector/SubInjector commands
Command/Property |
Range/Values |
Description |
AirGap |
Yes, No |
Allows using an air gap of 1 µl after the sample was drawn into the syringe. |
DispenseSpeed |
Slow, Middle, Fast |
Sets the injection speed. |
DrawSpeed |
Slow, Middle, Fast |
Sets the drawing speed. |
DwellTime |
0.0 s to 99.9 s |
Indicates the dwell time after the injection and before the needle is removed from the inlet. |
ErrorCode |
0 to 33 |
Indicates the error code of the sampler or injector if an error occurred. |
Inject |
Blank |
Performs an injection using the specified parameters. |
InjectMode |
SampleAir |
Sets the inject mode. Note: If no sampler is installed, not all injection modes are available. |
InjectWaitTime |
0.0 s to 10,000,000 s. |
Specifies the time between the inject command and the response of the autosampler. |
LoadsPerInject |
1 to 99 |
Sets the number of injections per sample. |
Position |
|
Sets the position of the sample in the rack. |
PreSolventWash |
0 to 99 |
Sets the number of pre-injection cleaning cycles during which the needle is washed with a solvent. Applies only when the selected injection mode is 0 (SampleAir). |
Pumps |
0 to 99 |
Determines how often the syringe plunger moves up and down before the sample is actually drawn. This eliminates gas bubbles, and thus enhances reproducibility. |
RackType |
Short Rack (1-6) |
Displays the rack type. |
Reset |
|
Resets the injector. |
SampleWash |
0 to 99 |
Sets the number of cleaning cycles with sample before injection. |
SolventSelect |
All A only B only C only |
(only available if a sampler is installed) Select the wash solvent. If no sampler is installed, you can select between using one or three wash vials (see Use3SolventVials). |
SolventWash |
0 to 99 |
Sets the number of post-injection cleaning cycles during which the needle is washed with a solvent. |
State |
On, Off |
Indicates the injection state (read-only): |
Status |
|
Indicates the autosampler's status (read-only). |
SyringeSpeed |
Slow, Fast |
Sets the syringe speed with which the syringe descends into the vial. |
SyringeHeightDown |
0 to 2 mm (for 1.5 ml vials) |
Sets how deep the needle descends into the vial. The Down parameter defines a position below the default Shimadzu position. |
SyringeHeightInj |
0 to 22 mm |
Defines how deep the needle descends into the injector during sample injection. |
SyringeHeightUp |
0 to 20 mm |
Sets how deep the needle descends into the vial. The Up parameter defines a position above the default Shimadzu position. |
Syringe |
0.5 µl, 5 µl, 10 µl, 50 µl, 250 µl |
Indicates the syringe type (read-only). |
Use3SolventVials |
Yes, No |
(only available if no sampler is installed) Sets the number of solvent vials used for washing (Yes = use 3 vials; No = use 1 vial). Also, see SolventSelect. |
VialType |
1.5 ml, 4 ml |
Indicates the vial type. |
Viscosity |
0.0 s to 99.9 s |
Sets the number of seconds that the pressure plunger of the syringe pauses after the pump and injection strokes. (Adapt this value to the viscosity of the sample.) While drawing sample before the injection, the system pauses for the defined time or a minimum of 4 seconds. The waiting time during the solvent wash is always 0.2 seconds. |
Volume |
0.1 µl to 8.0 µl |
Specifies the injection volume of the sample. |
VolumeForWashing |
60%, 80% |
Specifies the wash volume. |
WashSpeed |
Slow, Middle, Fast |
Sets the plunger speed for pumping or cleaning the syringe with sample liquid. |