Program Wizard/PGM Editor:
Sampler Parameter (Vanquish Autosampler)

Define the general settings for the Vanquish autosampler, such as speed and wash parameters.

Option

Description

 

 

Draw Speed

Specify the speed at which the sample is drawn into the needle.

 

 

Dispense Speed

Specify the speed at which sample is dispensed into the injection port.

 

 

Inject Wash Mode

Specify whether a wash cycle is to be performed for the needle:

  • NoWash: No wash cycle will be performed (recommended for typical applications).

  • BeforeDraw: Performs a wash step before the sample is drawn by the needle (the needle contains no sample liquid).

  • AfterDraw: Performs a wash step after sample liquid was drawn by the needle.

  • Both: Combines the BeforeDraw and AfterDraw options.

 

 

Wash Time

If a wash mode is defined, specify the duration of the wash cycle.

 

 

Wash Speed

Specify the flow speed of the wash liquid provided by the wash pump.

Note:

If you upgraded to Chromeleon 6.80 DU15a or later but use an PGM file created with a previous Chromeleon version, you will get an error message that the WashSpeed value is out of range, and the program cannot be started anymore. In order to fix your program, change the WashSpeed value manually: Divide the old value by 0.06 and enter the new value.  

 

 

Flow is delivered by

Specify the module delivering the flow to the inject valve.

 

 

Puncture Offset

The puncture offset defines how much the needle opens the septum. The carousel moves by the specified value while the needle remains in the septum, thus enlarging the hole in the septum and equalizing the pressure.

Specify the distance the carousel moves to enlarge the hole in the septum.

This can be useful in particular for vials that are filled to the maximum, or for poorly degassed samples.

 

 

Enable Stroke Synchronization

Select to enable pump stroke synchronization of the sampling and injection process with a Vanquish Quaternary pump or any other pump supporting pump stroke synchronization.

Note:

This option is only available if the VAS driver is linked to a corresponding pump. This can be done either in the Vanquish Autosampler Configuration Wizard or later on the Links tab in the sampler configuration.

 

Further Information