Formula for Amount Calculation (Rel. to
ISTD)
This function calculates the amount in an unknown
sample by means of the ratio of the area to the Internal
Standard. The prerequisite is that all peak areas of the sample have
been determined including the area of the internal standard.
p |
Indicates
the peak number. |
|
Indicates
the calibration function for the peak p that is calculated during
the calibration. For more information, refer to Creating
and Using Report Tables Calculating
the Peak Variable "Amount". |
|
Indicates
the determined area of peak p (for integration type = Area) |
Resp.Fact.p |
Indicates
the scaling factor: The scaling factor can be specific for each
peak; for example, to compensate the differing absorption behavior. |
ISTD |
Indicates
the internal standard. |
Dil.Fact. |
Corresponds
to the Dil.Fact. parameter from the sample list. |
Weight
(Sample Weight Factor) |
Is
a sample-specific factor that always refers to all peaks. |
n |
Indicates
the sample number |
The calculation of the calibration curve is
the precondition for amount calculation. For information about how the
calibration curve is calculated, refer to Theory
of Calibration Evaluation
with Various Standard Methods.