Pressure Limits
Depending on solvent composition and the selected Flow, pressures of up to 450 bar (= 45 MPa = 6525 psi) occur on the high-pressure side of a chromatography system. Chromeleon is capable of monitoring the pressure value provided by the pump. The user can define upper and lower pressure limits. If one of the limits is exceeded, Chromeleon stops the flow, displays an error message, and stops the sample batch, as necessary.
Also, refer to Pressure.Lower/Upper.Limits