Ensuring System Wellness (IC Devices)
Calibrating the AS Inject Port Volume
When to calibrate the inject port volume: |
After replacing the needle seal assembly, or if reproducibility between injections is poor. |
The inject port volume is the volume of tubing between the AS inject port and the injection valve. The autosampler uses this information to determine how much fluid to push through the line in order to position the sample correctly in the loop for precision injections.
Note:
For Sequential mode injections, the volume to be calibrated includes the needle seal line, the path through the diverter valve, and the diverter tubing to the injection valve.
Open the Inject Port Volume Calibration panel:
Open the Browser and open a panel tabset.
Select the Autosampler tab.
In the Wellness group, click the Calibrate IPTV button. The Inject Port Volume Calibration panel appears.
Follow the instructions on the panel to complete the calibration procedure. Refer to the appropriate notes below. The procedure is slightly different, depending on whether the AS is in Normal or Sequential mode.
Notes for Normal Mode:
Preparing: |
After disconnecting the needle seal line from the injection valve, route the line outside the autosampler or chromatography compartment. |
Step 1: |
Clicking Setup empties the needle seal line. The AS then dispenses 25 mL into the inject port. This nearly fills the needle seal line (minus a few microliters). |
Step 2: |
After clicking +1 uL, wait until the sound of the syringe motor stops before clicking +1 uL again. Continue clicking +1 uL, until a tiny drop of liquid appears at the end of the needle seal line. If you go too far and more than a single small drop of liquid appears, click -5 uL. The AS empties the line and then fills the line with 5 ml less than was cleared. You can then begin clicking +1 uL again. |
Step 3: |
After clicking Test, verify that a drop of liquid appears at the end of the needle seal line. If no drop appears, or if the size of the drop is different than previously observed, click -5 uL and repeat the procedure. |
Step 4: |
After clicking Done, the new calibration value is displayed on the panel. Write this value on the label inside the AS front door. |
Finishing: |
Reconnect the needle seal line to the injection valve. Tighten the fitting fingertight and then tighten an additional one-quarter turn using a 5/16-in wrench. |
Notes for Sequential Mode