ChemStation Translation Table
Translating ChemStation Methods

The ChemStation (CS) method contains parts of the Chromeleon PGM File, the QNT Method, and the Report Definition File (RDF). Chromeleon uses only the relevant information of a ChemStation method (i.e., Peak Table and Integration Events) to create a new QNT Method with the same name. A message appears if this method already exists. You can either overwrite the old method or cancel the import.

The table below lists the Integration Events of a ChemStation method and the equivalent detection parameters in Chromeleon:

 

ChemStation

Chromeleon

Comment

Slope Sensitivity

Sensitivity

Translation factor: 20

Peak Width

Peak Slice

Translation factor: 0.8

Area Reject

Minimum Area

Translation factor: 1/60

Height Reject

Minimum Height

 

Shoulders

Peak Shoulder Threshold

Off = 0ff, On = 0.1

Baseline Now

Lock Baseline

On

Baseline at Valleys

Valley to Valley

Off = Off, On = On

Baseline Hold

Lock Baseline

On = At Current Level, Off = Off

Tail Tangent Skim

-

 

Tangent Skim Mode

-

 

Area Sum

Peak Group Start

Peak Group End

On

Off

Integration

Inhibit Integration

Off = On, On = Off

Negative Peak

Detect Negative Peaks

Off = Off, On = On

Split Peak

-

 

Fixed Peak Width

Peak Slice

 

Auto Peak Width

-

 

Detect Shoulders

Peak Shoulder Threshold

Off = 0ff, On = 0.1

Shoulder Mode

-

-

Solvent Peak

-

 

Baseline Backwards

-

 

Baseline Next Valley

-

 

Unassigned Peaks

-

 

 

Chromeleon provides equivalents in the QNT Method for the following peak table parameters of a ChemStation method:

 

ChemStation

Chromeleon

Comment

Compound

Peak Name

 

RT

Ret. Time

 

RT Window Low

RT Window High

Window

Calculated from the difference: RT High - RT Low