Tab Dialog Box:
Pump Configuration (Gilson 331/332 and 333/334 Pumps)

Specify the GSIOC addresses of the pumps and indicate which head type is installed:

 

GSIOC:

Enter a separate GSIOC address for each connected pump. The address must match the address set on the instrument itself.

Each pump must have a unique GSIOC address. For the Gilson 331 and 333 pumps, enter the GSIOC address on the pump's control panel.

For stand-alone operation of the Gilson 331 or 333 pump, enter any address between 0 and 63. To operate the pump as master pump, the address must be 1.

For the Gilson 332 or 334 slave pump(s), the GSIOC address(es) must be 2 (and 3). Set the address via the switch on the pump's rear panel.

For more information about how to install the pumps, refer to Hardware Installation/Installing and Controlling Third-Party Devices:

 Gilson 331 and 332 Pumps

 Gilson 333 and 334 Pumps

 

 

Head type:

Select the installed head type(s).

For more information about the head types, refer to the respective pump manual.

 

 

Solvent valve

Select this check box for the pump if you use a Solvent Valve Block (SVB).